| Product Features | ||
| Medical & ITE safety approvals | ||
| 2MOPP input to output isolation | ||
| Leakage current≤ 100μA | ||
| Standby power ≤0.3W | ||
| 9V to 24V outputs, up to 48W | ||
| 3 types of AC inlet | ||
| Technical Data | ||
| Input Voltage Range: 90~264VAC | ||
| Input Current:1.1A at 90VAC | ||
| Output voltage Tolerance:±5% | ||
| Environmental Requirements | ||
| Operating Temperature: 0~40℃ | ||
| Storage Temperature: -20~60℃ | ||
| Operating Humidity:10~90%RH, non-condensing | ||
| Storage Humidity: 5~90%RH | ||
| Operating Altitude: 3,048 M | ||
| Safety Approvals | ||
| Safety aAgency/Mark | Medical | I.T.E |
| CB | IEC60601-1 | IEC60950-1 |
| UL | ANSI/AAMI ES60601-1 | - |
| CAN/CSA C22.2 NO.60601-1 | ||
| TUV-SUD/Mark | EN60601-1 | - |
| TUV-SUD/GS | - | EN60950-1 |
| CCC | - | GB4943.1(For Class I only) |
| PSE | - | J60950-1(For UE48-120300SPA1 only) |
| CE | - | EN60950 |
| FCC | - | FCC PART 15 |
| Mechanical Details | ||
| 121(L)*30.8(W)*50(H)mm | ||
| Applications: | ||
| Medical Device and ITE Equipment | ||
| Models & Ratings | |||||||
| Model No. | Voltage (V) | Current (A) | Ripple & Noise (Max: mV) PK-PK | Voltage tolerance | Line& Load Regulation | Efficiency (Average) Min | Start Up delay |
| 9.0-12.0 | 0.01-4.02 | 200 | ±5% | 88.62% | ≤3S | ||
| 12.1-13.0 | 0.01-3.70 | 200 | ±5% | 88.61% | ≤3S | ||
| 13.1-14.0 | 0.01-3.43 | 200 | ±5% | 88.60% | ≤3S | ||
| 14.1-15.0 | 0.01-3.20 | 200 | ±5% | 88.59% | ≤3S | ||
| 15.1-16.0 | 0.01-3.00 | 200 | ±5% | 88.59% | ≤3S | ||
| UE48-XXXYYYSPAZ | 16.1-17.0 | 0.01-2.83 | 200 | ±5% | Line:±1% | 88.61% | ≤3S |
| 17.1-18.0 | 0.01-2.67 | 200 | ±5% | Load: ±5% | 88.60% | ≤3S | |
| 18.1-19.0 | 0.01-2.53 | 200 | ±5% | 88.60% | ≤3S | ||
| 19.1-20.0 | 0.01-2.40 | 200 | ±5% | 88.59% | ≤3S | ||
| 20.1-21.0 | 0.01-2.29 | 200 | ±5% | 88.61% | ≤3S | ||
| 21.1-22.0 | 0.01-2.19 | 200 | ±5% | 88.62% | ≤3S | ||
| 22.1-23.0 | 0.01-2.09 | 200 | ±5% | 88.60% | ≤3S | ||
| 23.1-24.0 | 0.01-2.01 | 200 | ±5% | 88.62% | ≤3S |